Origin AFM Shut Down

  1. Click Stop to stop scanning the surface. Make sure it says “Engage” in the Sum and Deflection Panel. If not, click Withdraw.
  2. Make sure that the Video Monitor is open. If it is not open, Click on the camera icon on the bottom toolbar.
  3. Open the enclosure door.
  4. While watching the video monitor, raise the AFM head until there are a few millimeters of clearance between the specimen and cantilever.
  5. Using both hands, carefully place the AFM head upside down on the head stand.
  6. Press the release button on the head and gently lift the cantilever holder up, pulling the screw side up first and completely removing it from the AFM head.
  7. Mount the cantilever holder onto the loading stand, making sure the screws on the cantilever holder are facing toward the lever on the loading stand.
    1. Press down on the lever and align the back of the holder with the mounting pins. Press the back into place first and then the front.
    2. A lot of force does not need to be applied when loading it onto the stand.
  8. Using the screwdriver, loosen the top center screw on the cantilever holder.
  9. Remove the cantilever from the cantilever holder.
  10. Remove the specimen from the stage.
  11. Close and lock the Enclosure Door.
  12. Close the AFM Software.
  13. Turn OFF the AFM Controller.
  14. Turn OFF the Fiber Light.
  15. Sign out of the log book.